Pennsylvania Lab Startup, Relocation and Instrument Validation

Short answer: Opening or moving a clinical lab in Pennsylvania means settling the test menu, space plan and instruments before filing, because the PA DOH application asks for the premises plan, major equipment and quality control systems.
Each test then needs CLIA verification, or full validation if it is modified or laboratory developed, before patient testing.
Last reviewed: November 20, 2026. Primary sources: 28 Pa. Code 5.11, Permit, requirements, application and conditions, 28 Pa. Code 5.22, Responsibilities of owner and director, PA DOH, Instructions for the Application Process and Demographic Changes (REV 10/2025), 42 CFR 493.1253, Establishment and verification of performance specifications.
From plan to permit
A Pennsylvania permit application is more than a form. 28 Pa. Code 5.11(c) asks for the owner and director, the technical staff, a description and plan of the premises, a list of major equipment with manufacturer and model, the tests to be performed and the internal and external quality control systems (28 Pa. Code 5.11). That is why startup, space design and instrument selection should be settled before you file.
| Stage | What gets done | Rule it supports |
|---|---|---|
| Scope and test menu | Choose tests and CLIA complexity, map them to PA permit categories, decide on toxicology approval if needed | 28 Pa. Code 5.11(d); PA DOH instructions |
| Space design | Bench layout, specimen flow, storage, biosafety and the premises plan the application asks for | 28 Pa. Code 5.11(c)(3) |
| Instrument selection | Compare analyzers on menu, throughput, interfaces and QC; record manufacturer and model for the application | 28 Pa. Code 5.11(c)(4) |
| Applications | PA Clinical Laboratory Application with the CMS-116, director documents and the state filing fee | PA DOH instructions; CMS-116 |
| Verification and validation | Verify accuracy, precision, reportable range and reference intervals for unmodified FDA-cleared tests; establish full performance specifications for modified tests and LDTs | 42 CFR 493.1253 |
| Procedures, QC and PT | Written procedures for every test, in-date reagents, QC traceable to patient results, approved proficiency testing | PA DOH requirements for all labs |
| Survey | Certificate of Compliance or Accreditation labs get an on-site PA DOH survey before the permit is issued | PA DOH instructions |
Relocation. A move is a change of location, which the Bureau must be told about immediately on a Change of Status form; significant changes to physical facilities must be reported within 30 days (PA DOH instructions; 28 Pa. Code 5.22(a)). Plan the new space, re-verify instruments after they move, and keep the records ready for the next inspection. Method verification after relocation is covered in our method verification and validation guide.
What happens after you apply
| CLIA certificate | What PA DOH does after the application | Before testing |
|---|---|---|
| Certificate of Waiver | Issues the permit once documents are received and the CLIA fee is paid; may inspect as part of a waiver project | Permit issued |
| Provider-Performed Microscopy (PPM) | Issues the permit once documents and the CLIA fee are in; lab needs PT or twice-yearly peer review | Permit issued |
| Certificate of Compliance | Sends an acknowledgement; provisional approval to begin testing may be given; a Lab Examiner schedules a survey after about 90 days | Provisional approval possible |
| Certificate of Accreditation | Sends an acknowledgement; needs the accrediting organization's acknowledgement letter; schedules an on-site inspection | No testing until notified that licensure is in effect |
Source: PA DOH instructions.
Working on a Pennsylvania lab? Get a second set of eyes on your permit, director coverage or startup plan.
Verification and validation basics
Under 42 CFR 493.1253, an unmodified FDA-cleared or approved test must be verified for accuracy, precision, reportable range and reference intervals before patient testing. A modified test or a laboratory developed test needs performance specifications established for accuracy, precision, analytical sensitivity, analytical specificity including interfering substances, reportable range and reference intervals. See the SLC method verification and validation guide and LDT validation.
Working on a Pennsylvania lab? Get a second set of eyes on your permit, director coverage or startup plan.
Frequently asked questions
What does a Pennsylvania lab permit application need about the premises and equipment?
28 Pa. Code 5.11(c) asks for a description and plan of the premises, a list of major equipment with manufacturer and model, the tests to be performed and the internal and external quality control systems.
Do I need to tell PA DOH if my lab moves?
Yes. A change of location must be reported to the Bureau of Laboratories immediately on a Change of Status form, and significant physical facility changes within 30 days.
What is the difference between verification and validation?
For unmodified FDA-cleared or approved tests, CLIA requires the lab to verify accuracy, precision, reportable range and reference intervals. For modified tests and laboratory developed tests, the lab must establish all performance specifications, including analytical sensitivity and specificity (42 CFR 493.1253).
Can a new Certificate of Compliance lab start testing before the PA survey?
PA DOH may give provisional approval to begin testing after it acknowledges the application; a Lab Examiner schedules the survey after about 90 days. Accreditation-certificate labs may not test until notified that licensure is in effect.
Sources
- 28 Pa. Code 5.11, Permit, requirements, application and conditions
- 28 Pa. Code 5.22, Responsibilities of owner and director
- PA DOH, Instructions for the Application Process and Demographic Changes (REV 10/2025)
- 42 CFR 493.1253, Establishment and verification of performance specifications
Sound Lab Consultants is an independent clinical laboratory consulting firm headquartered at 25 Arizona Pl, Chandler, AZ 85225. We are not affiliated with or endorsed by the Pennsylvania Department of Health, CMS, CAP, COLA or The Joint Commission. This page is general information, not legal advice; confirm current requirements with the Bureau of Laboratories before filing.